INSTRUCTION MANUAL

Models H200 Single Channel K, J, T, E, R Thermometer

Table of Content Product Introduction 3
1-1 Features 3 1-2 Applications 3 Safety Information 4 2-1 Cautions 4 2-2 Safety Symbols 4 5 Operation 6 4-1 Unit Diagram 6 4-2 Measuring Procedure 7 4-3 Thermocouple Measuring Procedure 7 4-4 Offset Value Adjustment 7 4-5 Measuring Considerations 7 4-6 Data Hold Mode 7 4-7 Rec Mode 7 4-8 REL Mode 8 Auto Power Off Disable 8 RS232 PC Serial Interface Information 9 Battery Replacement 11 Service Information 12 Warranty Information 13

To see all available accessories and downloadable software, please visit anaheim sci entifi c. com

Product Introduction
1-1 Features Measures temperature fro m five different types of thermocouples Resolution Fast response ti me High level of accuracy Large LCD display Two year warranty
1-2 Applications Science Experiments HVAC Plant Maintenance Manufacturing Agriculture Quality Control

Safety Information Read the follow ing safety information care3fully before attempting to operate or
serv ice the meter. Only qualified personnel should perform repairs or serv icing not cov ered in this manual. 2-1 Cautions! NOT submerge the products mentioned in this manual in w ater or any other
ty pes of liquids. product is not designed for use in medical applications. The product can only
be used to measure body temperature simply for reference. 2-2 Safety Symbols

Dangerous, refer to this manual before using the meter. Conforms to requirements of European Union and European Fare Trade Association EFTA . Battery lev el is low. Don’t dispose this product as unsorted municipal waste. This instrument conforms to the follow ing standards EN61326 Electrical equipment for measurement, control and laboratory use. IEC61000-4-2 Electrostatic discharge immunity test. IEC61000-4-3 Radiated, radio-frequency, electromagnetic field immunity test. IEC61000-4-8 Power frequency magnetic field immunity test.

Specifications Unit of Measurement

Temperature Compensation

Thermocouple Types
°C or °F Automatic for cold junction on K & J type K, J, T, E & R

Thermocouple Type K Type J Type T Type E Type R

Resolution 0.1°C 0.1°F 0.1°C 0.1°F 0.1°C 0.1°F 0.1°C 0.1°F 1°C 1°F

Range
to 1300.0°C to -100.0°C to 2372.0°F
to -148.0°F to 1150.0°C to -100.0°C to 2102.0°F
to -148.0°F to 400.0°C to -100.0°C to 752.0°F
to -148.0°F to 900.0°C to -100.0°C to 1652.0°F
to -148.0°F 0 to 600°C
601 to 1700°C 32 to 1112°F
1113 to 3092°F

Ac c ur ac y *
+ % + 0.5°C + % + 1°C + % + 1°F
+ % + 1.8°F + % + 0.5°C + % + 1°C + % + 1°F
+ % + 1.8°F + % + 0.5°C + % + 1°C + % + 1°F
+ % + 1.8°F + % + 0.8°C + % + 1°C + % + 1.5°F
+ % + 1.8°F + 1 % + 5°C + % + 5°C + 1 % + 10°F
+ % + 10°F

Data Hold

Display freeze

Memory

Saves MAX. or MIN values with recall

Sampling Time

Approximately 1 second

Over Range Indicator