GXM60 Medical Single Output 60 Watt-Range Input High Efficiency Desk-Top Supply

FEATURES:
• Universal Input
• Small size, light weight
• Medically approved to UL2601-1, CSA No.
601.1-M90, EN60601-1, and CB certified to IEC601-1.
• Ultra efficient - Up to 90%
• Spill proof external enclosure
• 2 year warranty
• marked to LVD

SPECIFICATIONS:

AC INPUT 90-264 Vac, 47-63 Hz single phase. Input power is less than 75 W when output power is less than 60 W, 230 Vac in. Meets EN61000-3-2 harmonics standard. OUTPUT POWER Total continuous output power is 60 W 55 for any unit less than 8 V , 65 W peak for 60 s 10% duty cycle. Total power specified from supply output, derate for cable losses. OUTPUT REGULATION Supply regulation is +/- However, actual regulation limits are affected by cable losses on units without remote sense. Final regulation limits must include cable losses. Contact factory for assistance. INPUT CURRENT Maximum input current at minimum input voltage and output overload will be less than A. HOLD-UP TIME Output voltage stays within regulation for 20 ms from loss of ac input at full load, from 120 Vac input. MINIMUM LOAD No minimum load required. OVERLOAD PROTECTION Fully protected against short circuit and output overload. Short circuit protection is cycling type power limit and will automatically recover after removal of fault. OUTPUT NOISE rms, 1% pk-pk, 20 MHz bandwidth, differential mode. Measured with a scope probe directly across the output connector. Load terminated with a µF capacitor in parallel with a low Z 10 mF capacitor. TRANSIENT RESPONSE at supply is ms typical for return to regulation for a 50% load step, A/µs. Maximum voltage deviation is Cable resistance will effect maximum voltage deviation. OVER VOLTAGE PROTECTION Prevents output from producing excessive voltage in the event of a single fault failure. Voltage limited typically to 130% depending upon model. OVP firing reduces output to less than 50% of nominal voltage in less than 50 ms. OVERLOAD PROTECTION Factory set to begin power limiting at approximately 75 W.

EFFICIENCY
77 to 90% at full rated load depending upon model and ac line.

Units are typically > 85 % 230 Vac. Efficiency is measured at
the supply output, cabling losses are not considered and will
vary with different cables. Contact factory for assistance.

OVERSHOOT

Less than 3% overshoot at turn-on under nominal conditions.

TURN-ON TIME

Less than s at 115 Vac, 25 °C

INPUT PROTECTION

Both lines protected with internal ac fuses, provided on all units.

Designed to blow only if a catastrophic failure occurs in the unit

Fuse does not blow on short circuit or overload.

INRUSH CURRENT

Inrush is limited by internal thermistor. The inrush at 240 Vac,
averaged over the first ac half-cycle under cold start conditions
will not exceed 60 A.

EMI / EMC COMPLIANCE

All models include built-in EMI filtering.

Conducted Emissions

EN55011, Class B FCC Class B

Static Discharge

EN61000-4-2, 6 kV contact 8 kV air

RF Field Susceptibility

EN61000-4-3, 3V/meter

Fast Transients / Bursts

EN61000-4-4, 2 kV, 5 kHz

Surge Susceptibility

EN61000-4-5, 1 kV diff., 2 kV com.

Conducted RF Susceptibility EN61000-4-6, 3V

Voltage Sags & Surges

EN61000-4-11

MEDICAL SAFETY AGENCY

All models are approved to UL2601, CSA No. 601.1-M90,

IEC 601-1 1988 , EN 60601-1