ISOCON-3

ISOCON-3 Datasheet


ISOCON-3 USERS GUIDE

Part Datasheet
ISOCON-3 ISOCON-3 ISOCON-3 (pdf)
PDF Datasheet Preview
ISOCON-3 USERS GUIDE

ISOCON-3

MAINS POWERED ISOLATING SIGNAL CONVERTER

CAUTION:

This equipment is designed for connection to
mains voltages and must be used in accordance with this guide. If it is not, the safety protection
provided by the equipment may be impaired.

This equipment relies on double / reinforced insulation for safety and does not require a protective earth.

Whilst every effort has been taken to ensure the accuracy of this document, we accept no responsibility for damage, injury, loss or expense resulting from errors
or omissions, and reserve the right of amendment without notice.

IMO Industrial Interface

This document is issued by Industrial Interface Research Ltd and may not be reproduced in any way without the prior written permission of the company.

February 2002

IIG-060901

ISOCON-3 USERS GUIDE

CONTENTS

INTRODUCTION 3 Hardware Features 3

Isolation Details 3 CONNECTIONS 4 CONFIGURING THE 5

Voltage Input 6 Current Input 7 Millivolt mV Input 8 Potentiometer Input 9 Thermocouple 10 RTD Input 11 Output 12 CALIBRATING THE ISOCON 13 INSTALLATION 14 15 Incorrect Reading 15 Sensor Failure 15 SPECIFICATIONS 25°C 16

IMO Precision Controls Limited 1000 North Circular Road

Staples Corner, London NW2 7JP

Telephone +44 0 208 452 6444 Fax +44 0 208 450 274

Web Email:

IIG-060901

ISOCON-3 USERS GUIDE

INTRODUCTION

Hardware Features

The ISOCON-3 is a universal input Isolating Signal Converter. It can accept virtually every type of analogue input signal from millivolts to 40Vdc, mA, thermocouples, RTD’s etc. It also produces 3 types of analogue output voltage, mA source, or mA sink. The unit can be powered by any AC voltage between 90 and 264Vac at 50 or 60Hz. For DC voltage the ISOCON-6 is available which can be powered from any supply from 12 to 36Vdc and 12 to 32Vac.

The instrument is packaged in a very compact 12.5mm wide enclosure which can be mounted on standard TS35 DIN-rail.

The unit can also be equipped with 1 digital output which can be either a relay or an open collector output, or a second analogue output see DUALCON-3 . Note, units with above options are housed in a 17.5mm wide box. Isolation Details The ISOCON-3 has full 3 port isolation of 1000V between the Input Stage, Output Stage and Power Supply. The process input and output are double insulated internally from the power supply for safety reasons. Additionally the process input and output are isolated from each other for functional reasons.

UNPACKING

The instrument should be carefully inspected for signs of damage which may have occurred in transit. In the unlikely case that damage has been sustained, DO NOT use the instrument, as the safety insulation may have been compromised, but please retain all packaging for our inspection and contact your supplier immediately.

The instrument comes with the following items as standard 1 Isocon-3 Isolating Signal Converter 1 Isocon-3 User Guide

If the instrument has been factory configured the input and output details will be listed on the Serial number label on the side of the unit. If this label is blank then the unit will be set to its default configuration which is 4-20mA input and 420mA source output. Please check that the details on the side label are correct, especially the power supply voltage and frequency.

If re-configuration is required please refer to Section 4 of this manual.

IIG-060901

ISOCON-3 USERS GUIDE

CONNECTIONS

The ISOCON is housed in a compact DIN rail mounting enclosure, with 8 terminals, arranged in 4 rows of 2 terminals. Two rows are at the top of the front panel and 2 rows are at the bottom. All the sensor input terminals are on the bottom rows and the power supply and analogue outputs are on the top terminals.

The diagram below shows how to connect all the different input, output and power supply types.

PSU 21 OUTPUT
More datasheets: AM80A-048L-150F16 | AM80A-048L-240F10 | AM80A-300L-150F16 | AM80A-300L-065F33 | AM80A-048L-022F60 | B84112B0000K030 | B84112BL110 | TEL0023 | C3VFSC7 | AS1320 DB


Notice: we do not provide any warranties that information, datasheets, application notes, circuit diagrams, or software stored on this website are up-to-date or error free. The archived ISOCON-3 Datasheet file may be downloaded here without warranties.

Datasheet ID: ISOCON-3 520871