GSM7-24

GSM7-24 Datasheet


GSM7 Medical 7 Watt Single Output Global Performance Switchers

Part Datasheet
GSM7-24 GSM7-24 GSM7-24 (pdf)
Related Parts Information
GSM7-12 GSM7-12 GSM7-12
GSM7-15 GSM7-15 GSM7-15
GSM7-5 GSM7-5 GSM7-5
GSM7-28 GSM7-28 GSM7-28
PDF Datasheet Preview
GSM7 Medical 7 Watt Single Output Global Performance Switchers

FEATURES:
• World’s smallest medically approved 7 W supply
• x
• Wide-range ac input 85-264 Vac
• Medical Approval to UL2601-1 and EN60601-1
• EMI conducted to CISPR11 Class B
• Ideal for ultra-compact applications
• marked to LVD

SPECIFICATIONS:

Ac Input 85-264 Vac, 47-63 Hz single phase.

Holdup Time 16 ms minimum from loss of ac.

Output Power Normal continuous output power is 7 W, 9 W peak for 60 s maximum duration, 10% duty cycle. Factory set to begin power limiting at approximately 9 W.

Output Regulation from initial setpoint measured by changing load from no-load to 50% load or 50% load to full load in either direction. Initial sepoint tolerance is measured at 50% load. No minimum load is required to maintain proper regulation.

Overload Protection Fully protected against short circuit and output overload. Short circuit protection is cycling type power limit.

No Load Turn-on/Standby Full regulation at no load.

Output Noise rms, 1% pk-pk, 20 MHz bandwidth, differential mode. Measured with scope probe directly across output terminals of the power supply with load terminated with uF capacitor.

Voltage Adjustment Factory set with fixed resistors to maximize reliability.

Efficiency 72% minimum for the V model at full rated load, nominal input voltage. Efficiency increases as output voltage increases.

Input Protection Internal ac fuse provided on all units. Designed to open only if catastrophic failure occurs in the unit fuse does not blow on overload or short circuit. Internal 320 Vac MOV included. Units will not be damaged when tested for input transients in accordance with the following standards:

Level III Fast Level III

Environmental Designed for 0 to 50°C operation at full rated output power derate output current and total output power by per °C above 50°C. See Environmental specification on next page.

Medical EMI/EMC Compliance

All models include built-in EMI filtering to meet the
following emissions requirements:

EMI SPECIFICATIONS

COMPLIANCE LEVEL

Conducted Emissions

EN55011 Class B FCC Class B

Static Discharge

EN61000-4-2, 6 kV contact, 8 kV air

RF Field Susceptibility

EN61000-4-3, 3 V/meter

Fast Transients/Bursts

EN61000-4-4, 2 kV, 5 kHz

Surge Susceptibility

EN61000-4-5, 1 kV diff., 2 kV com.

Medical Safety Condor D.C. Power Supplies, Inc. declares under our sole responsibility that all GSM7 models are in conformity with the applicable requirements of EN60601-1 following the provisions of the Low Voltage Directive 73/23/EEC.

All GSM7 models are certified to be in compliance with the applicable requirements of UL2601-1 Patient Care Equipment , CSA No. 234 with additional tests to C22.2 No per T.I.L. CA-08 , EN Classification in accordance with sub-clause 5 of EN60601-1 Protection against electric shock = class II Degree of protection against electric shock = signal output or intermediate Protection against harmful ingress of water = ordinary no protection Not suitable for use in the presence of flammable anesthetic mixture Mode of operation = continuous

Earth Leakage Current Leakage current measured in ground wire connection to output common.

Model GSM7

Leakage Current* Leakage Current**
32 µA 55 µA
50 µA 85 µA

Test Voltage 132 Vac/60 Hz 264 Vac/50 Hz

Test Method UL2601-1 IEC601-1
* Normal condition **Single fault condition