GSM15 Medical
Part | Datasheet |
---|---|
![]() |
GSM15-12G (pdf) |
Related Parts | Information |
---|---|
![]() |
GSM15-12 |
PDF Datasheet Preview |
---|
GLOBAL PERFORMANCE SWITCHERS GSM15 Medical 15 Watt Global Performance Medical Switcher FEATURES: • Industry’s smallest 15 W medically approved switcher • Compact size x • Wide-range ac input 90-264 Vac • Less than 75 uA leakage current 120 Vac • Approved to UL2601-1, EN60601-1 • EMI to FCC, CISPR 11 Class B • Overvoltage protection standard • RoHS compliant models available G suffix • marked to LVD DC Power Supplies SPECIFICATIONS Ac Input 90-264 Vac, 47-63 Hz single phase. Class I or class II grounding. Input Current Maximum input current at 90 Vac, 60 Hz with full rated output load not to exceed A. Input Protection Internal ac fuse provided on all units. Designed to blow only if a catastrophic failure occurs in the unit -- Fuse does not blow on unsustained overload or short circuit. Inrush Current Inrush is limited by internal thermistors. The inrush at 240 Vac, averaged over the first ac half-cycle under cold start conditions will not exceed 37 A. Efficiency 69-85% depending on model. Overload Protection Fully protected against short circuit and output overload. Short circuit protection is cycling type power limit. Factory set to begin power limiting at 23 W. Overvoltage Protection Built in OVP on all models. Approximately 120-140% of output voltage. Output Noise rms, 1% Pk-Pk, 20 MHz Bandwidth, differential mode. Measured with noise probe directly across output terminals of the power supply. Transient Response Main Output - 500 µs max. response time for return to within of final value for a 50% load step change, A/us. Maximum voltage deviation is Hold-up Time 10 ms minimum from loss of ac input voltage at full load, nominal line 120 Vac . Temperature Coefficient / °C typical. EMI/EM Compliance All models include built-in EMI filtering to meet the following EMC requirements of IEC601-1-2. Performance Requirement EMC Standard Typical Margin Conducted Emissions EN55011, Class B FCC Class B 2 dB Class II Gnd 6 dB Class I Gnd Surge Discharge EN61000-4-2, Level 3 RF Field Susceptibility EN61000-4-3, Level 3 Fast Transients/Bursts EN61000-4-4, Level 3 500 V Surge Susceptibility EN61000-4-5, Level 3 500 V Conducted RF Susceptibility EN61000-4-6 Voltage Sags & Surges EN61000-4-11 Medical Safety Approvals All models are Certified to be in compliance with the applicable requirements of UL2601-1, IEC60601-1, CSA-C22.2 No. 601-1, EN60601-1. Leakage Current The maximum leakage current for GSM15 series will be as follows; 132Vac/60Hz UL2601-1 test method Connection Normal |
More datasheets: S553-0756-AE-F | S553-1006-AE-F | MGA-17516-BLKG | MGA-17516-TR1G | CYDD18S36V18-167BBXC | CYDD09S36V18-167BBXC | CYDD18S36V18-167BBXI | CYDD09S36V18-200BBXC | CYDD18S36V18-200BBXC | CYDD09S36V18-167BBXI |
Notice: we do not provide any warranties that information, datasheets, application notes, circuit diagrams, or software stored on this website are up-to-date or error free. The archived GSM15-12G Datasheet file may be downloaded here without warranties.