GNT30-48

GNT30-48 Datasheet


GNT30 MEDICAL SERIES SINGLE OUTPUT  POWER SUPPLY

Part Datasheet
GNT30-48 GNT30-48 GNT30-48 (pdf)
Related Parts Information
GNT30-24 GNT30-24 GNT30-24
GNT30-15G GNT30-15G GNT30-15G
GNT30-28 GNT30-28 GNT30-28
GNT30-24G GNT30-24G GNT30-24G
GNT30-28G GNT30-28G GNT30-28G
GNT30-5 GNT30-5 GNT30-5
GNT30-12 GNT30-12 GNT30-12
GNT30-48G GNT30-48G GNT30-48G
GNT30-15 GNT30-15 GNT30-15
GNT30-5G GNT30-5G GNT30-5G
GNT30-12G GNT30-12G GNT30-12G
PDF Datasheet Preview
GNT30 MEDICAL SERIES SINGLE OUTPUT  POWER SUPPLY

Low Power Medical Power Supplies 30 Watt AC/DC Universal Input Power Supply
• 90-264 Vac input range
• Overvoltage and short circuit protection
• Approved to UL, CSA and EN Standards
• Small 50 mm x 80 mm footprint
• Class II input available G2T30

The GNT30 Series is a 30 W universal input AC/DC Power supply in a very small footprint. With the medical approvals the GNT30 is ideal for a variety of medical device applications including, small single board computers, battery charging, and running small motors, pumps, and solenoids. The series with full approval to EN60601-1 Standard improves design-in time, and reduces end system compliance costs.

All specifications are typical at nominal input, full load at 25 OC unless otherwise stated

SPECIFICATIONS

OUPUT SPECIFICATIONS

Output Power
150 l.f.m. airflow

Total Regulation

Turn-on time

Transient response

A/ms 50% load deviation

Temperature Coefficient

Overvoltage protection

Short Circuit protection Power cycling
30 Watts max. See note 4 See table s max. max. dev. 1 ms recovery to 1% +/-0.03%/oC
125% +/-10% Yes

INPUT SPECIFICATIONS

Input Voltage range

Universal input

Input frequency Input surge current Ground Leakage current Input current

Input fuse
132 Vac 60 Hz 120 Vac 230 Vac F1,F2
100-240 nom 90 - 264 max Vac 55 Hz +/- 10 Hz 32 A max. 100 mA 180 mA single fault A

EMC INFORMATION

Conducted emissions EN55011

Radiated emissions EN55011

Line freq. harmonics EN61000-3-2

Voltage fluctuations EN61000-3-3

ESD Air

EN61000-4-2

ESD Contact

EN61000-4-2

Level B Level A Complies 8 kV contact 6 kV air

Radiated immunity Fast Transients EFT

Line surge immunity Conducted immunity Power freq. mag. field Voltage dip immunity

EN61000-4-3 EN61000-4-4

EN61000-4-5 EN61000-4-6 EN61000-4-8 EN61000-4-11
3 V/m 2 kV
1 kV diff /2 kV cm 3 Vrms 3 A/m 5 cycles 40 % vnom Watts.