GNT412ABTG

GNT412ABTG Datasheet


GNT400

Part Datasheet
GNT412ABTG GNT412ABTG GNT412ABTG (pdf)
Related Parts Information
GNT424BTG GNT424BTG GNT424BTG
GNT428ABG GNT428ABG GNT428ABG
GNT428CBEG GNT428CBEG GNT428CBEG
GNT428ABTG GNT428ABTG GNT428ABTG
GNT448ABTG GNT448ABTG GNT448ABTG
PDF Datasheet Preview
GNT400

Front End Series 400W AC/DC Universal Input Medical & ITE Power Supply
•Medical EN60601-1 and ITE EN60950 Certifications
•Compact 4” x 7” x U Channel 1U compatible
•100-240 VAC Input with Active PFC
•12 to 48 V DC Output at 400W 300W convection rating
•5 Vdc, 100 mA Standby Output
•12 Vdc Fan Output
•Approved to UL, CSA, and EN/IEC Standards
•Less than 100µA leakage
•Compliance to EN55011/22 Conducted Noise, Class B
•RoHS Compliant Category 6
2 Year Warranty

Specifications

All Specifications are typical at nominal input, full load at 25°C unless otherwise stated.

AC Input
100-240VAC, 50/60 Hz, 90-270VAC, 45-65 Hz max

Input Current
120VAC 4.2A. 230VAC 2.1A

Inrush Current

Active Circuit, 37A max. peak

Input Fuses

F1 and F2, 6.3A/250V for line and neutral

Power Factor
typical, minimum

Earth Leakage Current Output Power Peak Output Power
100µA at 120VAC, 60 Hz Normal 400µA at 264VAC, 50 Hz, SFC
400W continuous with 200lfm airflow See chart for convection ratings
450W for 1 minute, 10% duty cycle

Output Voltage

See chart

Voltage Adjustability Minimum Load Total Regulation Transient Response

Remote Sense
±5% minimum

Not required See chart
500µs to 1%, 50% load step, < 0.2A/µs Maximum voltage deviation is

Compensates for up to 500mV voltage drop

Turn On Time Hold-up Time Over/Undershoot Overload Protection Overvoltage Protection Ripple and Noise

ITE Safety Standards
500 ms max 20 ms at 400W
max Cycling type, 470W typical

See chart 1% pk-pk max., 20MHz BW 0.1µF and 10 µF caps on probe

CSA/EN/IEC/UL 60950-1

Medical Safety Standards EN/IEC/UL 60601-1, CSA22.2 No. 601-1

Thermal Considerations Keeping the following components below these temp. indicates sufficient cooling T5 90°C max., T1 100°C max Switching Frequency Output 170kHz ±8kHz, PFC 140kHz ±8kHz

MTBF
300K hours per Telcordia

Overtemperature Protection

Automatic

Efficiency
More datasheets: PI74FCT16244ATAE | PI74FCT16244ATVE | PI74FCT16244TAEX | PI74FCT16244CTVEX | PI74FCT16244ATAEX | PI74FCT162244ATA | AOZ8075DI | EM0712 | B39212B7835C710 | GNT424BTG


Notice: we do not provide any warranties that information, datasheets, application notes, circuit diagrams, or software stored on this website are up-to-date or error free. The archived GNT412ABTG Datasheet file may be downloaded here without warranties.

Datasheet ID: GNT412ABTG 506310