GLM75CG

GLM75CG Datasheet


GLC75 Commercial/GLM75 Medical

Part Datasheet
GLM75CG GLM75CG GLM75CG (pdf)
Related Parts Information
GLC75B GLC75B GLC75B
GLC75C GLC75C GLC75C
GLM75A GLM75A GLM75A
GLC75D GLC75D GLC75D
GLC75E GLC75E GLC75E
GLM75C GLM75C GLM75C
GLC75A GLC75A GLC75A
GLC75P GLC75P GLC75P
GLM75D GLM75D GLM75D
GLM75E GLM75E GLM75E
GLM75P GLM75P GLM75P
GLC75AG GLC75AG GLC75AG
GLC75FG GLC75FG GLC75FG
GLM75IG GLM75IG GLM75IG
GLC75BG GLC75BG GLC75BG
GLM75DG GLM75DG GLM75DG
GLM75EG GLM75EG GLM75EG
GLC75EG GLC75EG GLC75EG
PDF Datasheet Preview
GLOBAL PERFORMANCE SWITCHERS

GLC75 Commercial/GLM75 Medical
75 Watt Multiple Output Global Performance Switchers

FEATURES:
• Cost-effective multiple output power source
• Universal input 90-264 Vac
• 7.00”x 4.25”x Meets 1U height
• 2-year warranty
• Conducted EMI exceeds FCC Class B and CISPR 22

Class B Commercial models and CISPR 11 Class B Medical models
• Complies with EN61000-3-2 Class A
• Also available in single output versions
• Commercial UL1950 3rd Edition, CSA22.2 No. 950 and EN60950-1 approvals
• Medical Approved to UL2601-1, CSA22.2 No. 601.1-M90, and IEC/EN60601-1
• RoHS Compliant Model Available G suffix

SPECIFICATIONS

Ac Input 90-264 Vac, 47-63 Hz single phase.

Input Current Maximum input current at 120 Vac, 60 Hz with full rated output load not to exceed A.

Output Power Normal continuous output power is 75 W for unrestricted natural convection cooling or 110 W with 26 cfm airflow. During peak load conditions output regulation may exceed total regulation and noise limits.

Output Regulation Measured by ±40% load change from 60% rated load with all other outputs at 60% rated load and input voltage change from minimum to maximum ratings. Output #1 requires 1A minimum load for proper regulation of other outputs. Initial set tolerance is measured with all outputs at 60% of full rated load. Output #2 requires 0.5A minimum load for proper regulation.

Overload Protection Factory set to begin power limiting at approximately 120 W. Fully protected against short circuit and output overload. Short circuit protection is cycling type power limit.

Output Noise rms, 1% pk-pk, 20 MHz bandwidth, differential mode. Measured with noise probe directly across output terminals of the power supply.

Transient Response Main Output 500uSec typical response time for return to within of final value for a 50% load step change, Maximum voltage deviation is

Overvoltage Protection Built in on V1 with firing point set per table. OVP firing reduces output #1 and #2 to less than 50% of nominal voltage in 50 ms.

Voltage Adjust Factory set on standard unit however, optional potentiometer “-V” suffix adjusts voltage from V to OVP point V nominal on the +5 V output. Note Output #1 must not be more than 1% below nominal to achieve full output voltage range on Output regulation limits in some models may be exceeded when the main output is adjusted beyond +1% of nominal voltage. High voltage settings may degrade the reliability of the unit due to excessive power dissipation in some outputs.

Efficiency 68% -78% depending on model and load distribution.

Input Protection Internal ac fuse provided on all units. Designed to blow only if a catastrophic failure occurs in the unit.

Inrush Current Inrush limited by internal thermistors. Inrush at 240 Vac, averaged over the first ac half-cycle under cold start conditions will not exceed 37 A.

Hold Up Time 20 ms minimum from loss of ac input power at full load, nominal line 120 Vac .

Temperature Coefficient 0.03%/°C typical on all outputs.

Power Fail A standard TTL or CMOS compatible output goes low < V 5ms before output voltage drops more than 4% below nominal voltage upon loss of ac power. Signal is factory set to trip on 84 to 94 Vac brown-out depending upon incoming line impedance and distortion. Other settings are available through adjustment of built-in potentiometer consult factory for assistance . Output will stay low for 20 ms minimum.

EMI/EMC Compliance All models include built-in EMI filtering to meet the following emissions requirements:

EMI SPECIFICATIONS

Conducted Emissions GLC75 Conducted Emissions GLM75 Static Discharge RF Field Susceptibility Fast Transients/Bursts Surge Susceptibility Line Frequency Harmonics

COMPLIANCE LEVEL

EN55022 Class B FCC Class B EN55011 Class B FCC Class B EN61000-4-2, 6 kV contact, 8 kV air EN61000-4-3, 3 V/meter EN61000-4-4, 2 kV, 5 kHz EN61000-4-5, 1 kV diff., 2 kV com. EN61000-3-2 Class A

Commercial Safety All GLC models are approved to UL1950 3rd Edition, CSA22.2 No. 950, and EN60950-1.

Medical Leakage Current 70 uA 264 V 50 Hz normal conditions .

Medical Safety GLM models are approved to UL2601-1, CSA22.2 No. 601.1M90, IEC/EN60601-1. CB Report available.

All specifications are typical at nominal input, full load at 25OC unless otherwise stated

Commercial Medical RoHS Output

Model

Model Suffix* No.

Output

Output

Output

Output

Minimum Maximum A Maximum B

Output Peak

V1 OVP Set Noise P-P
More datasheets: GLM75C | GLC75A | GLC75P | GLM75D | GLM75E | GLM75P | GLC75AG | GLC75FG | GLM75IG | GLC75BG