MSM40A

MSM40A Datasheet


MSM40A/B/D Series

Part Datasheet
MSM40A MSM40A MSM40A (pdf)
PDF Datasheet Preview
MSM40A/B/D Series
40 Watt Triple Output Medical Open Frame Power Supply

High Efficiency

Wide Range AC Input

Fully Regulated DC Output

Lifetime Expectation >5 years

Time >20ms at full load

Outputs Floating

Safety Approval Class I

Elpac Part Number MSM40D MSM40B MSM40A

V1 I1 V2 I2 V3

I3 Total Regulation Typical Efficiency
5.1V 4.0A 24.0V 1.0A 0.4A ±5%
5.1V 4.0A 15.0V 2.0A 0.4A ±5%
5.1V 4.0A 12.0V 2.0A 0.4A ±5%

Input Voltage Input Frequency Input Current Inrush Current Zero Load Power Consumption Earth Leakage Current Typical
85 264VAC 100 240VAC Nominal 50 60Hz <1.3A rms <37A at 230VAC cold start <7.0W <25µA 132VAC 60Hz <40µA 264VAC 60Hz

Output

Output Voltage

See Table

Total Regulation
each output

Minimum Load

For V2 regulation, V1 current must be min 25% of V2, and <5x V2 current. For V3 regulation, min. load of 0.4A on V1

Delay
<750ms

Time
>20ms at full load, 115VAC

Ripple & Noise
<1% **

Over Voltage Protection

Over Temperature Protection Active Recoverable

Over Current Protetion
120 180%

Short Circuit Protection

Pulsing Mode Auto Recovery

General Efficiency MTBF Size Weight Power Density

Avg Efficiency 74% 230VAC min. 200,000 hours demonstrated lbs kg 2.0W/in3

Environmental Operating Temperature Storage Temperature Relative Humidity Cooling Vibration
0 50°C Full load to 50°C, derate linearly to 50% load at 70°C to +85°C Natural Convection All units production tested to 19.6m/s2

EMC & Safety Emissions Immunity Certified by UL to the following:
More datasheets: 614NGN | PF16.0W9 | ADNS-6090 | A2D250-AA06-84 | SBR12U45LH-13 | M52894 SL001 | M52894 SL005 | M52894 SL002 | ZXMP3F37DN8TA | 17000176A


Notice: we do not provide any warranties that information, datasheets, application notes, circuit diagrams, or software stored on this website are up-to-date or error free. The archived MSM40A Datasheet file may be downloaded here without warranties.

Datasheet ID: MSM40A 510994