Fluke type number Fluke 123 Fluke 124 Fluke 192B Fluke 196B, 196C, 196BM, 196CM Fluke 199B, 199C, 199BM, 199CM
Part | Datasheet |
---|---|
![]() |
196CM/003 (pdf) |
Related Parts | Information |
---|---|
![]() |
196BM/003 |
PDF Datasheet Preview |
---|
MRI adjustments with a Fluke 190 Series Application Note In modern medicine, Magnetic Resonance Imaging or “MRI” is one of the techniques used to study the inner structure of a living body on a non-invasive basis. The technique is an application of Nuclear Magnetic Resonance or NMR which is also used beyond the medical field to study the structure of matter. The main advantage of MRI over X-ray based techniques is in the better imaging capability of soft tissue, where X-ray techniques are more useful for study of the bone structure, for instance in case of fractures. Basics of MRI NMR is based on the fact that the atoms of many chemical elements and their isotopes have some degree of magnetic orientation. Under normal conditions, the individual magnetic fields of individual atoms show a random orientation, and the net field is zero Figure When placed in a very strong static magnetic field, the magnetic fields of the individual atoms line up, resulting in a net magnetization Figure The magnetic fields used to accomplish this are around to 4 tesla 1 T = 10000 gauss with a tendency to the ever stronger fields. Once the nuclei are aligned by this static magnetic field, an alternating magnetic field in a direction perpendicular to the main field will make the protons precess around the axis of their static orientation. When this occurs, the atoms turn out to have a resonance frequency in the RF-range. For instance the hydrogen atom 1H has a resonance frequency of approximately MHz when placed in a static magnetic field of 1 T. Given this resonance, the nuclei can be thought of as resonators that absorb energy from the applied RF-field. Nuclei of different elements and different isotopes all have different resonance frequencies. For as long as the RF-field is applied, all nuclei will resonate in phase, but once this stimulus ends the individual nuclei will lose synchronism, or de-phase, and return to their original position aligned with the static magnetic field. This process of returning to the original orientation takes a length of time that is again characteristic of each different type of nucleus. By measuring the resonance frequencies and the decay times, the individual elements present in the body can be determined. Figure 1 Normal randomly arranged magnetic orientation of nuclei. Figure 2 When placed in a strong magnetic field, the magnetic orientation of nuclei align with the external field. Oscilloscopes used in MRI For the proper functioning of the MRI-equipment, it is essential that the RF stimuli are switched on and off in a well controlled manner, and with good reproducibility. During development and adjustment of the equipment, a proper representation of these RF-bursts is most impor- From the Fluke Digital Library tant. For this purpose an oscilloscope is a most valuable piece of test equipment, if it can display the RF carrier frequency and is capable of showing the envelopes of the bursts properly. These envelopes show you everything that needs to be known about the process of activating and terminating the RF bursts. Many oscilloscopes, however, have difficulty working properly in the vicinity of the MRI-equipment because of the extremely high static magnetic fields. By nature, the CRT of the oscilloscope is sensitive to magnetic influences as it is based upon electrons travelling through the vacuum of the CRT. The Fluke ScopeMeters are all able to operate as compact, high performance oscilloscope that are well equipped to acquire and display electrical signals of various frequencies and that are not influenced by magnetic fields in contrast to a traditional CRT-based oscilloscope since the display used is an LCD. Fluke type number Fluke 123 Fluke 124 Fluke 192B Fluke 196B, 196C, 196BM, 196CM Fluke 199B, 199C, 199BM, 199CM Bandwidth 20 MHz 40 MHz 60 MHz 100 MHz 200 MHz Table 1 ScopeMeter models and their bandwidth. Max. sample rate single shot 25 MS/s 25 MS/s 500 MS/s 1 GS/s GS/s All Fluke ScopeMeters are equipped with a function called ‘Glitch Detect’. This function activates a digital overscan mechanism to find high-frequency signal content even when working at low speed timebase settings. An example of this is given in Figure At higher timebase speed, the number of cycles that is contained in an individual burst is quite limited, and it becomes more difficult to recognize the overall envelope of the signal see Figure Furthermore, at higher RF-carrier frequencies, the effects of aliasing may also influence the quality of the displayed bursts. Here, some of the additional functions found in the ScopeMeters come in handy to enhance the quality of the displayed waveform. Envelope function The Fluke 190B Series ScopeMeters are equipped with a socalled “Envelope”- function that builds the envelope over successive waveform-acquisitions on the display. See Figure For this, the extreme values that is the most positive and the most negative value for each of the horizontal positions over successive traces are stored, and these are overwritten only when newer values are found in a next trace that exceed the earlier ones at the same position. Figure 3 When placed in a static magnetic field B0 under the influence of an alternating field perpendicular to the static field, the nuclei start precessing. ScopeMeters used in MRI The Fluke ScopeMeters are available in different bandwidth and with different sample rates see table Given the RFfrequencies commonly used in MRI of and 127 MHz, the Fluke 190 Series ScopeMeters are the most appropriate instruments for test and alignment purposes. Figure 6 The envelope function of the Fluke 199B is used here with the same signal as in Figure Figure 4 At lower timebase setting, the ‘glitch detect’ function builds a solid envelope of a high frequency burst Fluke 199C . After just a few acquired traces, the envelop function shows a good representation of the overall envelope of the burst. On top of the envelope pattern, the most recently acquired waveform is then also displayed. Figure 5 A short duration burst is captured, but the envelope is more difficult to identify Fluke 199C . Digital persistence Next to the envelope function, the Fluke 190C Series of Color ScopeMeters also includes a socalled ‘Digital Persistence’ mode. In that mode, individual traces are stored in a dedicated display memory and each waveform fades away slowly. The decay time, or fade-out time, can be selected for short, medium or longer persistence, or it can be switched to have an ‘infinite persistence’. |
More datasheets: LTL-4261N | DCMY21WA4PA190 | DCMZ37PK87 | D20419-219 | D20419-216 | D20419-122 | D20419-220 | DBMMY25PE | MDM-25SH034F | 196BM/003 |
Notice: we do not provide any warranties that information, datasheets, application notes, circuit diagrams, or software stored on this website are up-to-date or error free. The archived 196CM/003 Datasheet file may be downloaded here without warranties.